Publikace: Fe3O4-modified thick film carbon-based amperometric oxidase-biosensor
Článekopen accesspeer-reviewedpublishedNačítá se...
Datum
Autoři
Waryo, Tesfaye T.
Begić, Sabina
Turkusić, Emir
Vytřas, Karel
Kalcher, Kurt
Název časopisu
ISSN časopisu
Název svazku
Nakladatel
Univerzita Pardubice
Abstrakt
An amperometric biosensor was developed from screen printed carbon electrodes modified with magnetite as the transducer element and Nafion-entrapped glucose-oxidase as the bio-recognition layer, cast by drop-evaporation. In a flowing stream of phosphate buffer (pH 7.5/0.05M), at a working potential of-125mV vs. Ag/AgCl with an, injection volume of 100 ul, the biosensor has linear range of 7-150 mg l-1 glucose, a sensitivity of 3.8 nA mg-1 1 cm2, and a precision of 2% at 10 mg l-1. Uric acid, xanthine and hypoxanthine at 10 and 1000 mg l-1 do not cause significant error relative to 100 mg l-1 glucose. Ascorbic acid at 100 mg l-1 causes 10-14% deviations- Dilute blood samples showed average recovery rate of 95%. Thus this model biosensor is precise, fairly senstitive and accurate.
Popis
Klíčová slova
biosenzory, uhlíkové elektrody, Amperometrie, tenké vrstvy