Laser induced breakdown spectrometry for quantitative analysis of Cd in artificial sweat leachate of jewellery
ČlánekOtevřený přístuppeer-reviewedDatum publikování
2017
Vedoucí práce
Oponent
Název časopisu
Název svazku
Vydavatel
University of Pardubice
Abstrakt
A new method has been developed for the determination of cadmium in artificial
sweat leachate using laser induced breakdown spectrometry (LIBS). Small volume
of liquid samples could be applied onto the solid carrier, dried under the IR lamp,
and subsequently analysed by the LIBS device. For calibration, two series of
standard solutions of artificial sweat were prepared: alkaline (pH 8) and acidic
(pH 5.5). The known volumes of the Cd standard (with concentration of 1 ± 0.001
g l–1) were added into the samples of artificial sweat; calibration standards
prepared ranging from 0-40 mg l–1 for acidic and 0-30 mg l–1 for alkaline sweat.
To optimize the LIBS measurements, the regression parameters (R2, AIC, MEP) were the criteria of choice. For our purposes, three cadmium lines at wavelengths
214.44 nm, 226.50 nm, and 228.80 nm were monitored. In model solutions, the
method developed has provided the limits of detection (LODs) of 0.08 mg l–1 for
acidic and of 0.06 mg l–1 for alkaline sweat.
The optimized method could then be used for the determination of Cd in
artificial sweat leachates of jewellery samples, when the values obtained were in
good agreement with the results from the reference analysis by ICP-OES. The
proposed method is fast, simple, and requiring only minimal volume of samples.
For analysis of leachate of jewellery, fabrics or products of daily consumption,
this approach employing LIBS can be an interesting alternative to commonly used
techniques, such as ICP OES or AAS.
Rozsah stran
p. 213–224
ISSN
1211-5541
Trvalý odkaz na tento záznam
Projekt
Zdrojový dokument
Scientific papers of the University of Pardubice. Series A, Faculty of Chemical Technology. 23/2017
Vydavatelská verze
Přístup k e-verzi
open access
Název akce
ISBN
978-80-7560-090-5