Direct determination of lead in bones using slurry sampling and high-resolution continuum source electrothermal atomic absorption spectrometry

ČlánekStatus neznámýpeer-reviewedThis is a postprint of an article published in Analytical Methods. The final authenticated version is available online at: https://doi.org/10.1039/c8ay02555j or https://pubs.rsc.org/en/content/articlelanding/2019/AY/C8AY02555J#!divAbstract.
Náhled

Datum publikování

2019

Vedoucí práce

Oponent

Název časopisu

Název svazku

Vydavatel

Royal Society of Chemistry

Abstrakt

A straightforward, quick, sensitive and reliable method is introduced for the determination of lead in bones using slurry sampling and high-resolution continuum source electrothermal atomic absorption spectrometry (HR-CS-ETAAS). The spectral interference caused by the molecular absorption of PO molecules with rotational fine structure coinciding with the analyte absorption at the most sensitive resonance line 217.001 nm in time was identified and successfully corrected by applying a mathematical correction algorithm using the spectrum obtained by vaporization of hydroxyapatite. The slurry preparation and measuring conditions were determined by means of a response surface methodology. Experiments were designed according to a 2(7-4) replicate (n = 3) fractional factorial design for seven factors (particle size, glycerol and HNO3 concentration, sonication time, concentration of the chemical modifier, pyrolysis and atomization temperature) each at three different levels, including central points. The optimized conditions were 100 mg of a ground sample with particle size < 315 mm, dilution in a liquid-phase composed of 10% w/w glycerol and 5.0% w/w nitric acid solutions, a sonication time of 2 min and the final slurry volume of 10.0 mL. A detection limit of 9.1 ug kg(-1) and characteristic mass of 7.6 pg were achieved using the suggested method under the optimized experimental conditions. Sufficient analyte stabilization was achieved by using 1 ug Pd and 50 ug citric acid. Accurate data were obtained with the use of matrix-free calibration. The accuracy of the method was established by analysing NIST SRM 1486 Bone Meal. Further, the results acquired for ten river otter samples by slurry sampling were compared with those determined after microwave-assisted digestion by inductively coupled plasma time of flight mass spectrometry (TOF-ICP-MS) to assess the accuracy of the method. The results obtained by the two methods were compared using a paired t-test (at the 95% confidence level) and showed no significant difference. The precision of the introduced method was better than 5.5%.

Rozsah stran

p. 1254-1263

ISSN

1759-9660

Trvalý odkaz na tento záznam

Projekt

Zdrojový dokument

Analytical Methods, volume 11, issue: 9

Vydavatelská verze

https://pubs.rsc.org/en/content/articlelanding/2019/AY/C8AY02555J#!divAbstract

Přístup k e-verzi

open access

Název akce

ISBN

Studijní obor

Studijní program

Signatura tištěné verze

Umístění tištěné verze

Přístup k tištěné verzi

Klíčová slova

Lead determination, Bone analysis, High-resolution continuum source AAS, Electrothermal atomic absorption spectrometry, Matrix effects, Design of experiments, Microwave-assisted digestion, Inductively coupled plasma mass spectrometry, Stanovení olova, Analýza kostí, HR-CS-ETAAS, ICP-MS, Analýza suspenzí, Interference matrice, Plánování experimentu, Frakcionální faktorové plánování