Practical experience with the ETV-ICP-MS prototype

ČlánekOtevřený přístuppeer-reviewedpublished
Náhled

Datum publikování

2019

Vedoucí práce

Oponent

Název časopisu

Název svazku

Vydavatel

University of Pardubice

Abstrakt

The aim of this study was to evaluate the commercial instrumental connection of the electrothermal vaporization unit (ETV) GF 5000 and the inductively coupled plasma orthogonal acceleration time-of-flight mass spectrometer (ICP-o-TOF-MS) Optimass 9500, both from the company GBC Scientific Equipment (Braeside, Australia). The prototype of the ETV unit was borrowed by the manufacturer for the purpose of its testing in the connection with the ICP-TOF-MS. Information on this technique instrumental arrangement has not been found in the literature yet. Therefore, the facts summarized in this paper are based mainly on our observations. The analytical performance of this technique achieved in our previous work is discussed and compared to competing spectrochemical methods. The attention was paid to the software and technical shortcomings observed within three years of working with this configuration. The influence of shortcomings of the instrument found during testing has been evaluated from the point of view of potential analytical applicability in its entirety and with respect to the economic impact on routine use.

Rozsah stran

p. 27 - 39

ISSN

1211-5541

Trvalý odkaz na tento záznam

Projekt

Zdrojový dokument

Scientific papers of the University of Pardubice. Series A, Faculty of Chemical Technology. 25/2019

Vydavatelská verze

Přístup k e-verzi

open access

Název akce

ISBN

978-80-7560-243-5

Studijní obor

Studijní program

Signatura tištěné verze

Umístění tištěné verze

Přístup k tištěné verzi

Klíčová slova

ETV-ICP-MS, user evaluation, software shortcomings, technical short-comings, switching valve, sample dosing monitoring