Publikace: Analytical Performance of Clay Paste Electrode and Graphene Paste Electrode-Comparative Study
Článekopen accesspeer-reviewedpublishedNačítá se...
Soubory
Datum
Autoři
Skowron, Ewelina
Spilarewicz-Stanek, Kaja
Guziejewski, Dariusz
Koszelska, Kamila
Metelka, Radovan
Smarzewska, Sylwia
Název časopisu
ISSN časopisu
Název svazku
Nakladatel
MDPI
Abstrakt
The analytical performance of the clay paste electrode and graphene paste electrode was compared using square wave voltammetry (SWV) and cyclic voltammetry (CV). The comparison was made on the basis of a paracetamol (PA) determination on both working electrodes. The influence of pH and SWV parameters was investigated. The linear concentration ranges were found to be 6.0 x 10(-7)-3.0 x 10(-5) and 2.0 x 10(-6)-8.0 x 10(-5) mol L-1 for clay paste electrode (ClPE) and graphene paste electrode (GrPE), respectively. The detection and quantification limits were calculated as 1.4 x 10(-7) and 4.7 x10(-7) mol L-1 for ClPE and 3.7 x 10(-7) and 1.2 x 10(-6) mol L-1 for GrPE, respectively. Developed methods were successfully applied to pharmaceutical formulations analyses. Scanning electron microscopy and energy-dispersive X-ray spectroscopy were used to characterize ClPE and GrPE surfaces. Clay composition was examined with wavelength dispersive X-ray (WDXRF).
Popis
Klíčová slova
square wave voltammetry, paracetamol, graphene, clay, carbon paste electrodes, sensors, square wave voltametrie, paracetamol, grafen, jíl, uhlíkové pastové elektrody, senzory