Digitální knihovna UPCE přechází na novou verzi. Omluvte prosím případné komplikace. / The UPCE Digital Library is migrating to a new version. We apologize for any inconvenience.

Publikace:
Possibilities of simultaneous voltammetric detection of tocopherols in non-aqueous media

Konferenční objektopen accesspeer-reviewedpostprint
Načítá se...
Náhled

Datum

Autoři

Jashari, Granit
Sýs, Milan
Metelka, Radovan
Švancara, Ivan

Název časopisu

ISSN časopisu

Název svazku

Nakladatel

BEST Servis

Výzkumné projekty

Organizační jednotky

Číslo časopisu

Abstrakt

An electrochemical study on possibilities of simultaneous detection of tocopherols based on their anodic oxidation at various electrode materials in non-aqueous supporting electrolytes was performed. Several working conditions, such as selection of working electrode, supporting electrolyte, effect of surfactant, water content and parameters of square-wave voltammetry, were optimized. An overlapping of corresponding peaks of alpha-, gamma- and delta- form. The separation of oxidation signals was improved at optimum working conditions, but the signals were not completely resolved. However, it was found that this phenomenon did not cause significant changes in peak currents for various concentration ratios of tocopherols, which could be possibly utilized in analytical applications.

Popis

Klíčová slova

Tocopherols, glassy carbon electrode, square wave voltammetry, food analysis, Tokoferoly, elektroda ze skelného uhlíku, square wave voltametrie, analýza potravin