Publikace: Electrocatalytic oxidation of NADH by poly-toluidine blue multiwall carbon nanotubes modified glassy carbon electrode
Článekopen accesspeer-reviewedpublishedNačítá se...
Soubory
Datum
Autoři
Vu Dai Long, /
Červenka, Libor
Název časopisu
ISSN časopisu
Název svazku
Nakladatel
University of Pardubice
Abstrakt
This study investigates a new approach to the amperometric determination of
NADH using a poly-toluidine blue multiwall carbon nanotubes modified glassy
carbon electrode (PTB-MWCNTs/GCE). The currents obtained from the
amperometric measurements at optimum conditions (phosphate buffer solution
(pH 7.0) containing 0.1 M KCl and constant applied potential: +150 mV vs.
Ag/AgCl/3.0 M KCl) were linearly correlated with the NADH concentrations. The
calibration curve was obtained for NADH concentrations in a range of 0.1-
100×10–6 mol l–1. The precision of the method was investigated by repeated
measurements (n = 10) of the 10 μmol NADH and the relative standard deviation
(RSD) was found to be 1.5 %. The limit of detection was found to be 0.04×10–6 mol
l–1 for amperometric method.