Publikace: Portable biosensor in analysis of surfactants
Konferenční objektopen accesspeer-reviewedpublished versionNačítá se...
Datum
Autoři
Khaled, Elmorsy
Hassan, Hassan
Abdelaziz, Mona
El-Attar, Rehab
Vytřas, Karel
Název časopisu
ISSN časopisu
Název svazku
Nakladatel
Univerzita Pardubice
Abstrakt
Sensitive potentiometric enzymatic biosensor was constructed and applied for analysis surfactants based on their inhibitory effect on acetylcholinesterase enzyme (AChE). The enzymatic activity was measured through monitoring hydrolysis of acetylcholine (Ach) with a disposable acetylcholine potentiometric sensor. Based on inhibition of AChE, different cationic, anionic and nonionic surfactants were determined in the concentration range from 0 to 40 µg mL-1 with detection limits reaching 0.07 µg mL-1 (depending on the nature of surfactants). The degree of AChE inhibition caused by different tested surfactants were as follows: cetylpyridinium chloride (CPC) > benzyldimethylhexadecyl ammonium chloride (BDHAC) > Hyamine (Hy) > cetyltrimethylammonium bromide (CTAB) > Triton X-100 (TX-100) > sodium dodecyl sulphate (SDS). The proposed method was applied for determination of surfactants in pharmaceutical formulation, detergents products and environmental samples.
Popis
Klíčová slova
biosensor, surfactant, potentiometry, biosenzor, tenzid, potenciometrie